Merck Magnetic perforated lid Microbial Air Sampler, for use with MAS-100 Sirius with 90 mm settle plates at 100 SLPM flow rate vanaf € 2.330,00 | css.shopping
Merck
Merck Magnetic perforated lid Microbial Air Sampler, for use with MAS-100 Sirius with 90 mm settle plates at 100 SLPM flow rate
Perforated lids are crucial for optimizing the physical and biological sampling efficiency of Andersen impaction-based microbial air samplers. Air is accelerated through the perforations in the lid, creating a jet that impacts the agar surface. The perforated lid A 90 mm for MAS-100 Sirius Microbial Air Samplers is designed to fit tightly on the instrument by easy and quick magnetic closure without the need for twisting. It is made from 316L stainless-steel which has good resistance towards a broad range of disinfectants including vaporized hydrogen peroxide and is suitable for autoclaving. The magnetic lid type A 90 mm has a sieve geometry of 300 holes with a diameter of 0.6 mm and is suitable for use with 90 mm agar plates. This sieve configuration delivers a nominal d50 value of 1.1 µm at a flow rate of 100 SLPM (standard liter per minute). Each lid is marked with laser-based marking without significantly altering the properties of the metal surface. The information oMeer tonenMinder tonen